Device and safety alerts families should see first.
When FDA or manufacturers publish CGM, pump, insulin, or related device notices, they appear here from the live Type1Cure safety feed — ahead of general news.
Medtronic MiniMed, Inc. · Oct 4, 2026
Medtronic MiniMed, Inc.: Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller en...
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposur... Action: On 08/20/2026, Technical Support called affected customers to inform that, MiniMed identified an issue in which certain MiniMed...
Open official noticeTandem Diabetes Care, Inc. · Sep 28, 2026
Tandem Diabetes Care, Inc.: Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2....
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P... Action: On 07/16/2026, an "URGENT MEDICAL DEVICE CORRECTION" communication was sent via certified mail to customers informing them that a...
Open official noticeMedtronic MiniMed, Inc. · Aug 17, 2026
Medtronic MiniMed, Inc.: MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis. Action: On 7/13/2026, field safety notices were sent to customers asking them to do the following: 1) Stop using the loaner pump and contact the firm to arrange for loaner pump return. 2) Contact your Healthcare Professional (H...
Open official noticeMedtronic MiniMed, Inc. · Aug 17, 2026
Medtronic MiniMed, Inc.: MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis. Action: On 7/13/2026, field safety notices were sent to customers asking them to do the following: 1) Stop using the loaner pump and contact the firm to arrange for loaner pump return. 2) Contact your Healthcare Professional (H...
Open official noticeMedtronic MiniMed, Inc. · Aug 17, 2026
Medtronic MiniMed, Inc.: MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis. Action: On 7/13/2026, field safety notices were sent to customers asking them to do the following: 1) Stop using the loaner pump and contact the firm to arrange for loaner pump return. 2) Contact your Healthcare Professional (H...
Open official noticeMedtronic MiniMed, Inc. · Jul 22, 2026
Medtronic MiniMed, Inc.: MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia. Action: On 6/16/2026 In-App Push Notifications were sent to customers entitled, Important Field Safety Corrective Action required on your MiniMed Go ap...
Open official noticeMedtronic MiniMed, Inc. · Jul 7, 2026
Medtronic MiniMed, Inc.: MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis... Action: On 5/27/2026, field safety notices were emailed to customers who were asked to do the following: 1) Update insulin pump software...
Open official noticeInsulet Corporation · Jun 29, 2026
Insulet Corporation: Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6...
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of in... Action: Insulet initially issued a press release on 05/26/2026 along with an "Urgent: Medical Device Correction" on the same day. Consig...
Open official noticeInsulet Corporation · Jun 29, 2026
Insulet Corporation: Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I...
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of in... Action: Insulet initially issued a press release on 05/26/2026 along with an "Urgent: Medical Device Correction" on the same day. Consig...
Open official noticeInsulet Corporation · Jun 29, 2026
Insulet Corporation: Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420...
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of in... Action: Insulet initially issued a press release on 05/26/2026 along with an "Urgent: Medical Device Correction" on the same day. Consig...
Open official noticeClamp Study · Nov 20, 2025
Clamp Study to Estimate the Relative Potency of GZR33 Versus Insulin Degludec at Steady State
RECRUITING, PHASE1. Interventions: GZR-33. Locations: Neuss, Germany | Neuss, Germany. This trial intends to investigate the pharmacodynamics, pharmacokinetics, safety, and tolerability of insulin GZR33 (hereafter referred to as GZR33) and estimate its potency in comparison with insulin degludec.
Open official notice